Eppendorf BioFlo 320 is vulnerable to due to VNC server...
Critical severity
Unreviewed
Published
May 26, 2026
to the GitHub Advisory Database
•
Updated May 26, 2026
Description
Published by the National Vulnerability Database
May 26, 2026
Published to the GitHub Advisory Database
May 26, 2026
Last updated
May 26, 2026
Eppendorf BioFlo 320 is vulnerable to due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted.
References